Table 2 Technical performance of the aggrecan capture OA-1 ARGS ELCL assay

From: Association between synovial fluid levels of aggrecan ARGS fragments and radiographic progression in knee osteoarthritis

Range of detection of standards

0.075 to 5 pmol ARGS/ml

Minimum required dilution of SF

1:4

Lower limit of detection in SF

0.3 pmol ARGS/ml SF

Parallelism (obtained/expected) of four SF samples diluted 1:4 to 1:4,000a

100% (78% to 120%)b

Spiking recovery of equimolar spiking of seven SF samples

116% (92% to 133%)b

Intra-assay CV, n = 14

3.5%

Inter-assay CV, n = 9

16.6%

  1. CV, coefficient of variation; ELCL, electrochemiluminescence; SF, synovial fluid. aSFs with 9, 95, 131, and 444 pmol ARGS/ml diluted and read within the range of detection of the standard curve. bMean (range).