From: Validation of the suction device Nimble for the assessment of skin fibrosis in systemic sclerosis

Fig. 1

Schemes of the working principles of the Nimble (a) and Cutometer (b) suction devices. The Nimble operates in a displacement-controlled fashion, with negative pressure drawing the skin into the probe opening until it reaches a defined height (h). The outcome measure is the pressure (pcl) needed for the specific tissue elevation. The Cutometer operates in a load-controlled fashion, with negative pressure drawing the skin into the probe opening until a maximum pressure is reached. The outcome measure is the elevation corresponding to a specific suction load. We extracted the maximum elevation (R0 in mm) for our study