Deep learning for detection of radiographic sacroiliitis: achieving expert-level performance

Abstract

Background

Radiographs of the sacroiliac joints are commonly used for the diagnosis and classification of axial spondyloarthritis. The aim of this study was to develop and validate an artificial neural network for the detection of definite radiographic sacroiliitis as a manifestation of axial spondyloarthritis (axSpA).

Methods

Conventional radiographs of the sacroiliac joints obtained in two independent studies of patients with axSpA were used. The first cohort comprised 1553 radiographs and was split into training (n = 1324) and validation (n = 229) sets. The second cohort comprised 458 radiographs and was used as an independent test dataset. All radiographs were assessed in a central reading session, and the final decision on the presence or absence of definite radiographic sacroiliitis was used as a reference. The performance of the neural network was evaluated by calculating areas under the receiver operating characteristic curves (AUCs) as well as sensitivity and specificity. Cohen’s kappa and the absolute agreement were used to assess the agreement between the neural network and the human readers.

Results

The neural network achieved an excellent performance in the detection of definite radiographic sacroiliitis with an AUC of 0.97 and 0.94 for the validation and test datasets, respectively. Sensitivity and specificity for the cut-off weighting both measurements equally were 88% and 95% for the validation and 92% and 81% for the test set. The Cohen’s kappa between the neural network and the reference judgements were 0.79 and 0.72 for the validation and test sets with an absolute agreement of 90% and 88%, respectively.

Conclusion

Deep artificial neural networks enable the accurate detection of definite radiographic sacroiliitis relevant for the diagnosis and classification of axSpA.

Background

Axial spondyloarthritis (axSpA) is a chronic inflammatory disease that mainly affects the axial skeleton, the sacroiliac joints and the spine. For many years, the detection of radiographic sacroiliitis has been the only way to make a definite diagnosis of the disease prior to the development of structural spinal damage. The presence of definite radiographic sacroiliitis (defined as sacroiliitis of at least grade 2 bilaterally or at least grade 3 unilaterally) is also a mandatory criterion of the modified New York criteria for ankylosing spondylitis (AS) [1]. Although magnetic resonance imaging (MRI) of the sacroiliac joints nowadays enables earlier diagnosis of axSpA, definite radiographic sacroiliitis can be detected at the time of diagnosis in about 33% of the patients with symptoms lasting up to 1 year and in about 50% of the patients with a symptom duration of 2 to 3 years [2]. Conventional radiography of the sacroiliac joints is therefore still recommended as the first imaging method in patients with suspected axSpA [3]. Furthermore, radiographic sacroiliitis — together with sacroiliitis on MRI — is included in the Assessment of Spondyloarthritis International Society (ASAS) classification criteria for axSpA [4]. Depending on the presence or absence of definite radiographic sacroiliitis, axSpA can be classified as either radiographic axSpA (r-axSpA, synonymous to AS) or non-radiographic axSpA (nr-axSpA) [5]. Such a classification could be relevant for both clinical practice (currently, the labels for biological disease-modifying antirheumatic drugs — bDMARDs are different for AS and nr-axSpA) and research (i.e., stratification or selection of patients in a clinical trial).

Although conventional radiography of the sacroiliac joints still plays an important role in both clinical practice and clinical trials, its reliability has been reported as mostly poor in a number of studies, even when assessed by expert readers [6,7,8,9,10]. In addition, it has been shown that untrained local readers perform worse than expert readers specialised in SpA [10]. One possible solution to achieve a comparable high accuracy as an expert in detecting radiographic sacroiliitis, even in non-specialised clinics, could be to develop an artificial intelligence-based model for the analysis of radiographs.

Deep learning has already produced remarkable results in the classification of medical and non-medical data. For example, deep neural networks have been trained to detect breast cancer in mammographs, to classify skin cancer or to label chest radiographs [11,12,13]. In all of these studies, the investigators did not develop a de novo model but applied a transfer learning approach using a pre-trained network. Such an approach allows the knowledge of pre-trained models from non-medical fields to be used for a new visual task, effectively reducing the amount of data required for training while increasing the accuracy of the models.

In the present study, we therefore aimed to develop and validate a deep neural network for the detection of definite radiographic sacroiliitis, using centrally scored images from two observational cohort studies.

Methods

Cohort description

In this project, we used imaging data from two independent sources: (1) Patients With Axial Spondyloarthritis: Multicountry Registry of Clinical Characteristics (PROOF) and (2) German Spondyloarthritis Inception Cohort (GESPIC).

PROOF is an ongoing study conducted in clinical practices in 29 countries and includes 2170 adult patients diagnosed with axSpA (non-radiographic or radiographic) ≤ 12 months before study enrolment and fulfilling the ASAS classification criteria for axSpA. In 1553 patients, radiographs of the sacroiliac joints were available for central reading.

GESPIC is a multicentre inception cohort study conducted in Germany and includes 525 patients with axSpA [14]. In 458 patients, radiographs of the sacroiliac joints were available for central reading.

Baseline characteristics of both cohorts are presented in Table 1.

Table 1 Baseline characteristics of patients with axial spondyloarthritis from the PROOF and GESPIC cohorts

Assessment of radiographic sacroiliitis

Radiographs of the sacroiliac joints were collected, digitised if necessary, anonymised and subsequently centrally graded by trained and calibrated readers using the modified New York criteria [1]:

Grade 0

Normal

Grade 1

Suspicious changes

Grade 2

Minimal abnormality: small localised areas with erosion or sclerosis, without alteration in the joint width

Grade 3

Unequivocal abnormality: moderate or advanced sacroiliitis with erosions, evidence of sclerosis, widening, narrowing or partial ankylosis

Grade 4

Severe abnormality: total ankylosis

In the PROOF study, images were first assessed by the local readers, then by central reader 1 (DP, board-certified rheumatologist with more than 10 years of experience in SpA imaging assessment), who was blinded to the results of the local assessment. In case of a disagreement on the presence of definite radiographic sacroiliitis (grade ≥ 2 bilaterally or grade ≥ 3 unilaterally) between the local and central reader 1, the radiograph was evaluated by central reader 2 (HH, board-certified rheumatologist with more than 10 years of experience in SpA imaging assessment), who was blinded to the previous assessments. The final decision on the presence of definite radiographic sacroiliitis and, therefore, on the classification as nr-axSpA or r-axSpA, was made based on the decision of two of the three readers.

In GESPIC, no local reading of radiographs was demanded; all collected images were scored independently by two trained and calibrated central readers (VRR and MT, board-certified rheumatologists with approximately 5 years of experience in SpA imaging assessment).

Image selection and pre-processing

The PROOF dataset consists of 1553 radiographs of the sacroiliac joints in DICOM (Digital Imaging and Communications in Medicine) format, varying in size, resolution and quality (Fig. 1). The Horos Project DICOM Viewer (version 4.0.0, www.horosproject.org) was used to adjust the greyscale levels of all images and to convert them to the Tagged Image File Format (TIFF) afterwards. Images including other body parts such as the thoracic spine were manually cropped to the pelvis. The final dataset for building the model was split randomly into training (1324 radiographs, 85%) and validation datasets (229 radiographs, 15%).

Fig. 1
figure 1

Flowchart for the selection of cases from the PROOF (training and validation set) and GESPIC (test set) studies

For testing the generalisability across datasets, we defined two subgroups in the GESPIC dataset: (1) patients with the presence or absence of definite radiographic sacroiliitis in the opinion of two readers (n = 352) and (2) patients with disagreement of the central readers on the final classification (n = 106). Images in the test dataset were pre-processed exactly as the training and validation datasets.

Model training

Model training was performed on a dedicated Ubuntu 18.04 workstation with two Nvidia GeForce RTX 2080ti graphic cards as well as on a GPU node of the Berlin Institute of Health (BIH) high-performance computing cluster using four Nvidia Tesla V100 graphic cards. All model training was mainly performed using Python (version 3.7) including the fastAI application programming interface, which is built on top of PyTorch [15, 16].

As a base model, we used a convolutional neural network (ResNet-50 architecture) pre-trained on the ImageNet-1k dataset, which includes over 1.28 million images [17]. The images were augmented prior to training through various transformations including flipping, rotation of up to 10°, magnification of up to 1.1, lighting variations and warping. We further utilised the mix-up method during training, originally introduced by Zhang et al. [18], in which images of different classes (nr-axSpA and r-axSpA) are combined during training to reduce memorisation of noisy labels and increase overall model robustness. As a loss function, we used cross entropy label smoothing, which reduced high-confidence predictions of the models, thus supporting regularisation and avoiding overfitting with subsequent improved generalisation of the models on new data (e.g., test dataset). The optimal learning rate for training was determined using a learning rate range test [19]. Model training was performed with cyclical [19], discriminative learning rates (as initially implemented by Howard and Ruder [20]) and a progressive re-sizing approach, starting with image sizes of 224 × 224 pixels (which is the default input size for the ImageNet pre-trained ResNet-50) and next increasing the resolution to 512 × 512 pixels and then to 768 × 768 pixels. During training, first only the last two classification layers of the model were trained, with the weights of the other network layers remaining frozen. A total of 100 epochs were trained, monitoring the area under the receiver operating characteristics curve (AUC) on the validation dataset and saving the model weights on every improvement. After 100 epochs, the weights of the model with the highest AUC value were re-loaded, the model was unfrozen and again trained for another 100 epochs (training all layers of the network), while monitoring the AUC and saving the weights at every improvement. This approach was repeated for all image resolutions. The size of the mini batches was 64 for 224 × 224 pixels, 32 for 512 × 512 pixels and 84 for 768 × 768 pixels. The training for lower resolutions could be performed at our local workstation, while for 768 × 768 pixels, computation has been performed on the HPC for Research cluster of the Berlin Institute of Health. Overall, model training took approximately 24 h on our local machine and an additional 6 h on the cluster. After training, Gradient-weighted Class Activation Mapping (Grad-CAM) was used to create activation maps for verification that the model actually used the sacroiliac joints to determine if definite radiographic sacroiliitis was present [21].

Statistical analysis

Statistical analysis was performed using the “R” statistical environment (version 3.6), the “tidyverse”, “ROCR” and “irr” libraries [22,23,24,25]. Raw predictions of the model on the validation dataset as well as on the test dataset using an image resolution of 768 × 768 pixels were exported from the python environment as comma-separated values and imported into “R”. ROC curves and precision-recall curves were plotted, and the AUC was calculated. Three different cut-offs were chosen through repeated cross validation for the calculation of sensitivity and specificity, the first cut-off favouring sensitivity, the second favouring specificity and the third aiming at balancing both. Confusion matrices were constructed using the predefined cut-offs. Cohen’s kappa and the percentage absolute agreement were used to assess the agreement between the human readers and the network. Ninety-five per cent confidence intervals for calculated kappa values were estimated using bootstrapping with 1000 repetitions. A p-value of < 0.05 was considered statistically significant.

Ethics approval

Both PROOF and GESPIC were approved by the local ethics committees of each study centre in accordance with the local laws and regulations and were conducted in accordance with the Declaration of Helsinki and Good Clinical Practice. The institutional review board of the Charité — Universitätsmedizin Berlin additionally approved GESPIC. Written informed consent was obtained from all patients.

Results

Definite radiographic sacroiliitis in the opinion of two readers was present in 873 (65.9%) patients from the training set (PROOF, n = 1324) and in 150 (65.5%) patients from the validation set (PROOF, n = 229). In a total of 369 (27.9%) and 63 (27.5%) patients in the training and validation sets, respectively, there was a discrepancy between the local reader and central reader 1, which automatically resulted in the involvement of central reader 2. A total of 146 (11.0%) and 37 (16.2%) patients in the training and validation sets were re-classified after the central reading, meaning that, in these cases, the ratings of both central readers differed from the rating of the local reader.

In the test set (GESPIC), both readers agreed on the presence of radiographic sacroiliitis in 223 (48.7%) cases and on the absence of radiographic sacroiliitis in 129 (28.2%) cases and disagreed in 106 (23.1%) cases.

Model performance in the validation dataset

There was excellent performance of the model on the validation dataset. The receiver operating characteristics curve (ROC) analysis showed an AUC of 0.969. For the precision-recall (PR) curve, an average AUC of 0.989 was achieved. Both the local and central expert readers remained below the ROC and PR curves and were therefore outperformed by the accuracy of the model. We propose three cut-offs to convert the floating-point predictions into integer values with 1 representing the presence of definite radiographic sacroiliitis and 0 its absence. Cut-offs weighting sensitivity over specificity and specificity over sensitivity were used in order to find the optimal balance between both parameters (defined as the maximum sum between sensitivity and specificity). The first cut-off value, which favours sensitivity over specificity, was calculated to be 0.475, resulting in a sensitivity of 0.993 and a specificity of 0.177 for the detection of r-axSpA. The second cut-off, which favoured specificity over sensitivity, was 0.787, resulting in a sensitivity of 0.753 and a specificity of 0.987. The third cut-off was 0.724, resulting in a sensitivity of 0.880 and a specificity of 0.949. ROC curves and precision-recall curves of model performance are shown in Fig. 2a, and Table 2 summarises performance results as confusion matrices with kappa values and values of absolute agreement.

Fig. 2
figure 2

Receiver operation characteristics curve and precision-recall curve for the model performance in detecting definite radiographic sacroiliitis (classified as non-radiographic or radiographic axial spondyloarthritis) on the validation (a) and test (b) dataset as well as the corresponding area under the curve and average precision. Individual values for the local and the central expert reader are displayed as a triangle or dot in a. Since the reference standard in b was the agreement of two independent readers, their accuracy is not presented

Table 2 Confusion matrices for the three proposed cut-offs for the model predictions regarding the presence of definite radiographic sacroiliitis on the validation dataset

Model performance on the independent dataset

The model’s performance on the test dataset was assessed in two subsets. In the first subset, which comprised the cases where the two readers agreed on either the presence or absence of definite radiographic sacroiliitis (n = 352), the model performed slightly worse than on the validation dataset with an AUC value of 0.936 and an average precision (AP) value of 0.962. Again, we applied the three cut-offs as calculated from the validation dataset: The first cut-off, which weights sensitivity over specificity, yielded a sensitivity of 0.982 and a specificity of 0.264. For the second cut-off, which weights specificity over sensitivity, a sensitivity of 0.816 and a specificity of 0.930 were achieved. For the third cut-off, aiming at defining optimal performance in terms of both performance measures, we calculated a sensitivity of 0.915 and specificity of 0.806. Figure 2b shows the ROC- and precision-recall curves for the model performance on the test dataset. Figure 3 demonstrates the different values for sensitivity and specificity achieved for different cut-offs on the test and validation datasets. Table 3 provides confusion matrices for the three proposed cut-offs and the overall accuracy. Figure 4 shows examples of Grad-CAM maps of the neural network for predictions on the test datasets.

Fig. 3
figure 3

Sensitivity and 1-specificity (false positive rate) on the test and validation datasets using different cut-off values for the model predictions regarding the presence of definite radiographic sacroiliitis (classification as non-radiographic or radiographic axial spondyloarthritis). We analysed three cut-off values, indicated by vertical dashed lines. Cut-off 1 weights sensitivity over specificity, cut-off 2 weights specificity over sensitivity and cut-off 3 aims to be the optimal balance between the two performance measures. Cut-offs were only calculated on the validation dataset and then applied to the test and validation datasets

Table 3 Confusion matrices for the three proposed cut-offs for the model predictions regarding presence of definite radiographic sacroiliitis on the test dataset
Fig. 4
figure 4

Gradient-weighted Class Activation Maps (Grad-CAMs) of the model for selected images taken from the test dataset. Grad-CAMs provide visual explanations for the model decision, as they highlight the important image regions that led to the model’s decision. The examples presented show that the model almost exclusively and correctly focusses on the sacroiliac joints to predict whether definite radiographic sacroiliitis is present or not

For the second subset of cases, with disagreement between the two central readers (n = 106), the algorithm resulted in the following classification distribution: cut-off 1 (favouring sensitivity), 8 nr-axSpA and 98 r-axSpA; cut-off 2 (favouring specificity), 73 nr-axSpA and 33 r-axSpA; and cut-off 3 (balanced sensitivity and specificity), 53 nr-axSpA and 53 r-axSpA.

The interrater agreement between the human readers, as measured by Cohen’s kappa on the entire test dataset (n = 458), was moderate with k = 0.53 (95% CI 0.46–0.61) and a percentage agreement of 76.9%. The agreement between reader 1 and the neural network was similar with k = 0.54 (95% CI 0.46–0.62) and a percentage agreement of 77.3%. The agreement between reader 2 and the neural network was slightly better with k = 0.57 (95% CI 0.49–0.65) and a percentage agreement of 80.3%.

Discussion

In this study, we successfully developed and tested an artificial intelligence model for the detection of radiographic sacroiliitis on conventional radiographs. With this model, we achieved an excellent model accuracy on the validation data. Furthermore, we demonstrated the generalisability of our model on a test dataset of novel data, achieving a performance at least comparable to that of two human experts.

Although magnetic resonance imaging is increasingly used for the detection of sacroiliitis in industrialised nations, radiographs are still important. In many countries, radiographs remain the first and often the only imaging procedure for examining patients with axSpA because MRI is expensive and not widely available. The detection of definite radiographic sacroiliitis is important for both the diagnosis and classification of axSpA. At the same time, it is well known that conventional radiographs are not very reliable in detecting sacroiliitis [6,7,8,9,10]. In the present study, we used a large and unique dataset to train, validate and test the model. The resulting performance was at least as good as (but most likely better than) the performance of an experienced reader with expertise in radiographic sacroiliitis assessment. The neural network was able to achieve almost the same level of performance in both the validation and training sets, indicating a high level of reliability and robustness of the model. Our model can therefore be used as an additional diagnostic aid in clinical practice and as a classification tool in research projects involving patients with axSpA.

Neural networks have already been applied to a variety of medical imaging data, including radiographs but, to our knowledge, not for the detection of spondyloarthritis [11,12,13, 26]. However, a low generalisability, i.e., poor performance of the models on new data, is an important challenge in training neural networks. A new meta-analysis on ‘deep learning performance against healthcare professionals’ by Kim et al. revealed methodological shortcomings that are present in many published studies on deep learning in medicine [27]. They criticised that many studies either did not compare the performance of their model with that of a human domain expert or assessed the performance of their model on a different dataset than the one used for human performance assessment, resulting in excessively high accuracies, mainly due to over-adaptation, which consequently have a low generalisability [27]. Similar observations were made by Yao et al., who showed that, while they identified 155 studies on deep learning in medicine, the studies often lacked external validation data [28]. However, the use of external validation data is an important measure to prove generalisability. It has been shown that medical computer vision models adapt poorly to the use of different scanners or imaging protocols, and the lack of external validation is likely to result in poor generalisability of the model to new data [29]. In a recent study, McKinney et al. evaluated the performance of a neural network for the detection of breast cancer in mammographs, showing that the network surpassed human performance [11]. They used different datasets from different studies to train and test their developed models and were thus able to demonstrate sufficient generalisability of their models.

Similar to their approach, we also used a heterogeneous training dataset with radiographs from different imaging sites and achieved a good generalisability of the developed model, with the performance on the test data being only slightly inferior. In our study, the test data were independent from the validation data in terms of both patients and readers [18, 30, 31]. While the heterogeneity of our training dataset already reduced the risk of overfitting on systematic image noise, e.g., to device-specific image features, we further increased generalisability by applying progressive re-sizing and the integration of mix-up as well as label smoothing into model training.

Our study has some limitations. First, the reference for the training of the model was the judgement of a limited number of human readers (2 or, in the case of discrepancy in the PROOF study, 3). Although both central readers in the PROOF study had many years of experience in the reading of radiographs of the sacroiliac joints, the complex sacroiliac joint anatomy and heterogeneity of radiographic techniques and quality have introduced some uncertainty into the final classification used as a reference. In the independent dataset, we selected primarily only cases where both readers agreed to be the reference standard for the evaluation of the model. This approach was chosen because we believe that these cases are most likely to be true positive or true negative, while in the cases with a discrepancy, the truth is not known. Nonetheless, the neural network-based classification of the discrepant cases was well balanced with the balanced cut-off indicating that our algorithm is also applicable in such rather difficult cases. It is noteworthy that, despite all the uncertainty related to the assessment of radiographic sacroiliitis, a high level of agreement between the neural network’s judgement and the human consensus judgement was achieved in both validation and test datasets.

Another limitation is related to the chosen sets — all patients were diagnosed with axSpA. The performance of the algorithm in patients with undiagnosed back pain and suspected axSpA in the diagnostic setting is not known and should be investigated in future studies.

Conclusions

Radiographs of the sacroiliac joints are commonly used for the diagnosis and classification of axial spondyloarthritis, but the reliability of the definite radiographic sacroiliitis detection is usually low. Convolutional neural networks can detect radiographic sacroiliitis on pelvic radiographs with at least the same level of accuracy as a human expert. Utilisation of the proposed computer vision model could thus enable highly accurate detection of definite radiographic sacroiliitis, even in non-specialised sites.

Availability of data and materials

The data that support the findings of this study are available from the corresponding author, JLV, upon reasonable request.

Abbreviations

AP:

Average precision

AS:

Ankylosing Spondylitis

ASAS:

Assessment of Spondyloarthritis International Society

AUC:

Area under the curve

axSpA:

Axial spondyloarthritis

DICOM:

Digital Imaging and Communications in Medicine

GESPIC:

German Spondyloarthritis Inception Cohort

Grad-CAM:

Gradient-weighted Class Activation Mapping

MRI:

Magnetic resonance imaging

nr-axSpA:

Mon-radiographic axial spondyloarthritis

PROOF:

Patients With Axial Spondyloarthritis: Multicountry Registry of Clinical Characteristics

r-axSpA:

Radiographic axial spondyloarthritis

ROC:

Receiver operating characteristics curve

PR:

Precision-recall

TIFF:

Tagged Image File Format

References

  1. van der Linden S, Valkenburg HA, Cats A. Evaluation of diagnostic criteria for ankylosing spondylitis. A proposal for modification of the New York criteria. Arthritis Rheum. 1984;27(4):361–8. https://doi.org/10.1002/art.1780270401.

    Article  PubMed  Google Scholar 

  2. Poddubnyy D, Brandt H, Vahldiek J, Spiller I, Song IH, Rudwaleit M, Sieper J. The frequency of non-radiographic axial spondyloarthritis in relation to symptom duration in patients referred because of chronic back pain: results from the Berlin Early Spondyloarthritis Clinic. Ann Rheum Dis. 2012;71(12):1998–2001. https://doi.org/10.1136/annrheumdis-2012-201945.

    Article  PubMed  Google Scholar 

  3. Mandl P, Navarro-Compan V, Terslev L, et al. Eular recommendations for the use of imaging in the diagnosis and management of spondyloarthritis in clinical practice. Ann Rheum Dis. 2015;74(7):1327–39. https://doi.org/10.1136/annrheumdis-2014-206971.

    Article  CAS  PubMed  Google Scholar 

  4. Rudwaleit M, van der Heijde D, Landewe R, Listing J, Akkoc N, Brandt J, Braun J, Chou CT, Collantes-Estevez E, Dougados M, Huang F, Gu J, Khan MA, Kirazli Y, Maksymowych WP, Mielants H, Sorensen IJ, Ozgocmen S, Roussou E, Valle-Onate R, Weber U, Wei J, Sieper J. The development of assessment of Spondyloarthritis International Society classification criteria for axial Spondyloarthritis (part II): validation and final selection. Ann Rheum Dis. 2009;68(6):777–83. https://doi.org/10.1136/ard.2009.108233.

    Article  CAS  PubMed  Google Scholar 

  5. Boel A, Molto A, van der Heijde D, Ciurea A, Dougados M, Gensler LS, Santos MJ, de Miguel E, Poddubnyy D, Rudwaleit M, van Tubergen A, van Gaalen FA, Ramiro S. Do patients with axial Spondyloarthritis with radiographic sacroiliitis fulfil both the modified New York criteria and the ASAS axial spondyloarthritis criteria? Results from eight cohorts. Ann Rheum Dis. 2019;78(11):1545–9. https://doi.org/10.1136/annrheumdis-2019-215707.

    Article  PubMed  Google Scholar 

  6. Spoorenberg A, de Vlam K, van der Linden S, et al. Radiological scoring methods in ankylosing spondylitis. Reliability and change over 1 and 2 years. J Rheumatol. 2004;31(1):125–32.

    PubMed  Google Scholar 

  7. Christiansen AA, Hendricks O, Kuettel D, Hørslev-Petersen K, Jurik AG, Nielsen S, Rufibach K, Loft AG, Pedersen SJ, Hermansen LT, Østergaard M, Arnbak B, Manniche C, Weber U. Limited reliability of radiographic assessment of sacroiliac joints in patients with suspected early spondyloarthritis. J Rheumatol. 2017;44(1):70–7. https://doi.org/10.3899/jrheum.160079.

    Article  PubMed  Google Scholar 

  8. Yazici H, Turunc M, Ozdoğan H, et al. Observer variation in grading sacroiliac radiographs might be a cause of 'sacroiliitis’ reported in certain disease states. Ann Rheum Dis. 1987;46(2):139–45. https://doi.org/10.1136/ard.46.2.139.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Poddubnyy D, Rudwaleit M, Haibel H, Listing J, Marker-Hermann E, Zeidler H, Braun J, Sieper J. Rates and predictors of radiographic sacroiliitis progression over 2 years in patients with axial spondyloarthritis. Ann Rheum Dis. 2011;70(8):1369–74. https://doi.org/10.1136/ard.2010.145995.

    Article  PubMed  Google Scholar 

  10. Van den Berg R, Lenczner G, Feydy A, et al. Agreement between clinical practice and trained central reading in reading of sacroiliac joints on plain pelvic radiographs: results from the Desir cohort. Arthritis Rheum. 2014;66(9):2403–11. https://doi.org/10.1002/art.38738.

    Article  Google Scholar 

  11. McKinney SM, Sieniek M, Godbole V, et al. International evaluation of an Ai system for breast cancer screening. Nature. 2020;577(7788):89–94. https://doi.org/10.1038/s41586-019-1799-6.

    Article  CAS  PubMed  Google Scholar 

  12. Esteva A, Kuprel B, Novoa RA, Ko J, Swetter SM, Blau HM, Thrun S. Dermatologist-level classification of skin cancer with deep neural networks. Nature. 2017;542(7639):115–8. https://doi.org/10.1038/nature21056.

    Article  CAS  PubMed  Google Scholar 

  13. Irvin J, Rajpurkar P, Ko M, et al. Chexpert: a large chest radiograph dataset with uncertainty labels and expert comparison. Thirty-Third Aaai Conference on Artificial Intelligence / Thirty-First Innovative Applications of Artificial Intelligence Conference / Ninth Aaai Symposium on Educational Advances in Artificial Intelligence 2019:590–597.

  14. Rudwaleit M, Haibel H, Baraliakos X, Listing J, Märker-Hermann E, Zeidler H, Braun J, Sieper J. The early disease stage in axial spondylarthritis: results from the German spondyloarthritis inception cohort. Arthritis Rheum. 2009;60(3):717–27. https://doi.org/10.1002/art.24483.

    Article  CAS  PubMed  Google Scholar 

  15. Paszke A, Gross S, Massa F, Lerer A, Bradbury J, Chanan G, Killeen T. Lin Z, Gimelshein N, Antiga L, Desmaison A, Kopf A, Yang E, DeVito Z, Raison M, Tejani A, Chilamkurthy S, Steiner B, Fang L, Bai J, Chintala S. PyTorch: An Imperative Style, High-Performance Deep Learning Library. In;  Wallach H, Larochelle H, Beygelzimer A,  d\textquotesingle Alch\'{e}-Buc F, Fox E, Garnett R, editors. Advances in Neural Information Processing Systems. Red Hook: Curran Associates, Inc. 2019;32. https://proceedings.neurips.cc/paper/2019/file/bdbca288fee7f92f2bfa9f7012727740-Paper.pdf.

  16. Howard J, Gugger S. Fastai: a layered API for deep learning. Information. 2020;11(2):108. https://doi.org/10.3390/info11020108.

    Article  Google Scholar 

  17. He K, Zhang X, Ren S, Sun J. Deep residual learning for image recognition. The IEEE Conference on Computer Vision and Pattern Recognition (CVPR); 2016.

    Google Scholar 

  18. Zhang H, Cisse M, Dauphin YN, Lopez-Paz D. Mixup: beyond empirical risk minimization. International Conference on Learning Representations; 2017.

    Google Scholar 

  19. Smith LN. Cyclical learning rates for training neural networks. 2017 IEEE Winter Conference on Applications of Computer Vision (WACV); 2017:464-72. IEEE. https://doi.org/10.1109/WACV.2017.58.

  20. Howard J, Ruder S. Universal language model fine-tuning for text classification. Proceedings of the 56th Annual Meeting of the Association for Computational Linguistics (Volume 1: Long Papers). Melbourne: Association for Computational Linguistics; 2018:328–39. https://www.aclweb.org/anthology/P18-1031https://doi.org/10.18653/v1/P18-1031.

  21. Selvaraju RR, Cogswell M, Das A, et al. Grad-Cam: visual explanations from deep networks via gradient-based localization. Proceedings of the IEEE international conference on computer vision; 2017.

    Google Scholar 

  22. R Core Team. R: a language and environment for statistical computing. Vienna: R Foundation for Statistical Computing; 2013.

    Google Scholar 

  23. Wickham H, Averick M, Bryan J, Chang W, McGowan L, François R, Grolemund G, Hayes A, Henry L, Hester J, Kuhn M, Pedersen T, Miller E, Bache S, Müller K, Ooms J, Robinson D, Seidel D, Spinu V, Takahashi K, Vaughan D, Wilke C, Woo K, Yutani H. Welcome to the Tidyverse. J Open Source Softw. 2019;4(43):1686. https://doi.org/10.21105/joss.01686.

    Article  Google Scholar 

  24. Sing T, Sander O, Beerenwinkel N, Lengauer T. Rocr: visualizing classifier performance in R. Bioinformatics. 2005;21(20):3940–1. https://doi.org/10.1093/bioinformatics/bti623.

    Article  CAS  PubMed  Google Scholar 

  25. Gamer M, Lemon J, Fellows I, Singh P. Irr: various coefficients of interrater reliability and agreement. R package version 0.84.1; 2010.

  26. Liu X, Faes L, Kale AU, Wagner SK, Fu DJ, Bruynseels A, Mahendiran T, Moraes G, Shamdas M, Kern C, Ledsam JR, Schmid MK, Balaskas K, Topol EJ, Bachmann LM, Keane PA, Denniston AK. A comparison of deep learning performance against health-care professionals in detecting diseases from medical imaging: a systematic review and meta-analysis. Lancet Digit Health. 2019;1(6):e271–e97. https://doi.org/10.1016/S2589-7500(19)30123-2.

    Article  PubMed  Google Scholar 

  27. Kim DW, Jang HY, Kim KW, Shin Y, Park SH. Design characteristics of studies reporting the performance of artificial intelligence algorithms for diagnostic analysis of medical images: results from recently published papers. Korean J Radiol. 2019;20(3):405–10. https://doi.org/10.3348/kjr.2019.0025.

    Article  PubMed  PubMed Central  Google Scholar 

  28. Yao AD, Cheng DL, Pan I, Kitamura F. Deep learning in neuroradiology: a systematic review of current algorithms and approaches for the new wave of imaging technology. Radiology. 2020;2(2):e190026.

    Google Scholar 

  29. Shin H-C, Roth HR, Gao M, Lu L, Xu Z, Nogues I, Yao J, Mollura D, Summers RM. Deep convolutional neural networks for computer-aided detection: CNN architectures, dataset characteristics and transfer learning. IEEE Trans Med Imaging. 2016;35(5):1285–98. https://doi.org/10.1109/TMI.2016.2528162.

    Article  PubMed  Google Scholar 

  30. He T, Zhang Z, Zhang H, et al. Bag of tricks for image classification with convolutional neural networks. 2019 IEEE/CVF Conference on Computer Vision and Pattern Recognition (CVPR); 2019. p. 558–67.

    Google Scholar 

  31. Szegedy C, Vanhoucke V, Ioffe S, Shlens J, Wojna Z. Rethinking the inception architecture for computer vision. 2016 IEEE Conference on Computer Vision and Pattern Recognition (CVPR); 2016. p. 2818–26.

    Google Scholar 

Download references

Acknowledgements

LA is grateful for her participation in the BIH Charité Junior Clinician and Clinician Scientist Program and KKB is grateful for his participation in the BIH Charité Digital Clinician Scientist Program all funded by the Charité — Universitätsmedizin Berlin, the Berlin Institute of Health, and the Deutsche Forschungsgemeinschaft. We also thank the BIH for providing access to their HPC for Research cluster for our analysis. The authors thank Bettina Herwig for language editing.

Funding

GESPIC was initially supported by the German Federal Ministry of Education and Research (Bundesministerium für Bildung und Forschung, BMBF). As scheduled, BMBF was reduced in 2005 and stopped in 2007; thereafter, complementary financial support was obtained also from Abbott, Amgen, Centocor, Schering–Plough, and Wyeth. Starting in 2010, the core GESPIC cohort was supported by AbbVie. The PROOF study is funded by AbbVie. We thank AbbVie for allowing us to use the PROOF dataset for the aim of the current study. Open Access funding enabled and organized by Projekt DEAL.

Author information

Authors and Affiliations

Authors

Contributions

KKB and JLV collected and pre-processed imaging data, trained the computer vision model, interpreted the data and drafted the manuscript. LA interpreted the data and substantially revised the manuscript. SMN contributed to study design and substantially revised the manuscript. HH, VRR, MT, MP, FP and JR acquired and analysed study data. JS, MR, BH, MRM and KGH contributed to conception of the study and substantially revised the manuscript. DP collected imaging data, interpreted the data, contributed to study design and drafted the manuscript. All authors contributed to writing and correcting the manuscript and have approved the final version.

Corresponding author

Correspondence to Janis L. Vahldiek.

Ethics declarations

Ethics approval and consent to participate

Both PROOF and GESPIC cohorts were approved by the local ethics committees of each study centre in accordance with the local laws and regulations and is being conducted in accordance with the Declaration of Helsinki and Good Clinical Practice. GESPIC was additionally approved by a central ethics committee of the coordinating centre. Written informed consent to participate was obtained from all patients.

Consent for publication

Not applicable.

Competing interests

K.K.B., L.C.A., J.R., M.T. and M.M. have nothing to disclose.

J.L.V. reports non-financial support from Bayer, non-financial support from Guerbet, non-financial support from Medtronic, and personal fees and non-financial support from Merit Medical, outside the submitted work.

S.M.N. reports grants from the German Research Foundation (Deutsche Forschungsgemeinschaft), personal fees from Bracco Imaging, personal fees from Canon Medical Systems, personal fees from Guerbet, personal fees from Teleflex Medical GmbH, and personal fees from Bayer Vital GmbH, outside the submitted work.

H.H. reports personal fees from Pfizer, personal fees from Janssen, personal fees from Novartis, personal fees from Roche, and personal fees from MSD, outside the submitted work.

V.R.R. reports personal fees from Abbvie and personal fees from Novartis, outside the submitted work.

M.P. reports personal fees from Novartis and personal fees from AbbVie, outside the submitted work.

F.P. reports personal fees from AbbVie, personal fees from AMGEN, personal fees from BMS, personal fees from Celgene and from MSD, grants and personal fees from Novartis, personal fees from Pfizer and from Roche, and personal fees from UCB, outside the submitted work.

J.S. reports grants from AbbVie, during the conduct of the study; personal fees from AbbVie, personal fees from Novartis; personal fees from Pfizer; personal fees from Roche; personal fees from UCB; personal fees from Boehringer Ingelheim; personal fees from Janssen; and personal fees from Merk, outside the submitted work.

M.R. received honoraria and/or consulting fees from AbbVie, BMS, Celgene, Janssen, Eli Lilly, MSD, Novartis, Pfizer, Roche, and UCB Pharma.

B.H. reports grants from Abbot, grants from Actelion Pharmaceuticals, grants from Bayer Schering Pharma, grants from Bayer Vital, grants from BRACCO Group, grants from Bristol-Myers Squibb, grants from Charite Research Organisation GmbH, grants from Deutsche Krebshilfe, grants from Essex Pharma, grants from Guerbet, grants from INC Research, grants from lnSightec Ud, grants from IPSEN Pharma, grants from Kendlel MorphoSys AG, grants from Lilly GmbH, grants from MeVis Medical Solutions AG, grants from Nexus Oncology, grants from Novartis, grants from Parexel Clinical Research Organisation Service, grants from Pfizer GmbH, grants from Philipps, grants from Sanofis-Aventis, grants from Siemens, grants from Teruma Medical Corporation, grants from Toshiba, grants from Zukunftsfond Berlin, grants from Amgen, grants from AO Foundation, grants from BARD, grants from BBraun, grants from Boehring Ingelheimer, grants from Brainsgate, grants from CELLACT Pharma, grants from CeloNova Bio-Sciences, grants from GlaxoSmithKline, grants from Jansen, grants from Roehe, grants from Sehumaeher GmbH, grants from Medronic, grants from Pluristem, grants from Quintiles, grants from Roehe, grants from Astellas, grants from Chiltern, grants from Respicardia, grants from TEVA, grants from Abbvie, grants from AstraZenaca, and grants from Galmed Research and Development Ltd., outside the submitted work.

K.G.H. reports personal fees from AbbVie, personal fees from Pfizer, personal fees from MSD, and personal fees from Roche, outside the submitted work.

D.P. reports grants and personal fees from AbbVie, during the conduct of the study; grants and personal fees from AbbVie; personal fees from BMS; personal fees from Celgene; grants and personal fees from Lilly; grants and personal fees from MSD; grants and personal fees from Novartis; grants and personal fees from Pfizer; personal fees from Roche; and personal fees from UCB, outside the submitted work.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bressem, K.K., Vahldiek, J.L., Adams, L. et al. Deep learning for detection of radiographic sacroiliitis: achieving expert-level performance. Arthritis Res Ther 23, 106 (2021). https://doi.org/10.1186/s13075-021-02484-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13075-021-02484-0

Keywords