Accuracy and tolerability of self-sampling of capillary blood for analysis of inflammation and autoantibodies in rheumatoid arthritis patients—results from a randomized controlled trial

Abstract

Background

Rheumatoid arthritis (RA) requires early diagnosis and tight surveillance of disease activity. Remote self-collection of blood for the analysis of inflammation markers and autoantibodies could improve the monitoring of RA and facilitate the identification of individuals at-risk for RA.

Objective

Randomized, controlled trial to evaluate the accuracy, feasibility, and acceptability of an upper arm self-sampling device (UA) and finger prick-test (FP) to measure capillary blood from RA patients for C-reactive protein (CRP) levels and the presence of IgM rheumatoid factor (RF IgM) and anti-cyclic citrullinated protein antibodies (anti-CCP IgG).

Methods

RA patients were randomly assigned in a 1:1 ratio to self-collection of capillary blood via UA or FP. Venous blood sampling (VBS) was performed as a gold standard in both groups to assess the concordance of CRP levels as well as RF IgM and CCP IgG. General acceptability and pain during sampling were measured and compared between UA, FP, and VBS. The number of attempts for successful sampling, requests for assistance, volume, and duration of sample collection were also assessed.

Results

Fifty seropositive RA patients were included. 49/50 (98%) patients were able to successfully collect capillary blood. The overall agreement between capillary and venous analyses for CRP (0.992), CCP IgG (0.984), and RF IgM (0.994) were good. In both groups, 4/25 (16%) needed a second attempt and 8/25 (32%) in the UA and 7/25 (28%) in the FP group requested assistance. Mean pain scores for capillary self-sampling (1.7/10 ± 1.1 (UA) and 1.9/10 ± 1.9 (FP)) were significantly lower on a numeric rating scale compared to venous blood collection (UA: 2.8/10 ± 1.7; FP: 2.1 ± 2.0) (p=0.003). UA patients were more likely to promote the use of capillary blood sampling (net promoter score: +28% vs. −20% for FP) and were more willing to perform blood collection at home (60% vs. 32% for FP).

Conclusions

These data show that self-sampling is accurate and feasible within one attempt by the majority of patients without assistance, allowing tight monitoring of RA disease activity as well as identifying individuals at-risk for RA. RA patients seem to prefer upper arm-based self-sampling to traditional finger pricking.

Trial registration

DRKS.de Identifier: DRKS00023526. Registered on November 6, 2020.

Introduction

Rheumatoid arthritis (RA) is one of the most common autoimmune diseases [1] causing chronic inflammation and loss of function. Since RA is a chronic disease, patients require a life-long care with close monitoring of disease activity. However, as there is an increasing shortage of specialized health care for patients with rheumatic and musculoskeletal diseases [2], patients often face challenges of receiving timely and adequate on-site care. This shortage has lately been exacerbated by the COVID-19 pandemic, in which a temporary reduction in appointment slots has been observed [3].

In addition to the need for remote monitoring of disease activity in patients with RA, it is also important to detect the disease at an early stage, i.e., identifying patients at-risk for developing RA, since delayed diagnosis is associated with joint damage, loss of function, and lower treatment efficacy [4]. Being “at risk of RA” is closely related to the presence of antibodies against cyclic citrullinated peptides (anti-CCP) or rheumatoid factor (RF IgM)). At present, however, assessment of the presence of RA-related autoantibodies is exclusively done by healthcare professionals via venipuncture.

Remote monitoring of disease activity may substantially improve clinical care [5]. A challenge for comprehensive monitoring of diseases, such as RA is the blood collection that ideally should be carried out anywhere and anytime [5, 6] so that no clinical visit and consultation of a health care provider is required. Blood sample collection needs to be easy to perform, painless, and provide samples of sufficient quality and quantity for reliable and rapid analysis. Successful implementation of such practice can potentially improve early detection and monitoring of disease [7], save labor and costs in clinical trials and practice [8,9,10], and enable flexible drug level monitoring [11]. Furthermore, self-laboratory results could be made available for clinical visits, thereby reducing potential delays in treatment decisions.

The most commonly used self-sampling method of the blood is the withdrawal of capillary blood from fingertips, which is perceived as painful and often provides insufficient sample quantities [12] for assessment of inflammation and autoantibodies. New blood sampling devices may have the potential to overcome these hurdles and optimize remote monitoring [12, 13]. We therefore tested the accuracy, feasibility, and acceptability of self-sampling of inflammation markers and autoantibodies in RA patients, comparing two forms of self-sampling of the capillary blood (upper arm and finger tip) with standard venous blood collection by health care professionals.

Methods

Study design and randomization

This study was a prospective, single-center, cross-sectional, parallel, two-group, non-blinded, randomized controlled trial (WHO International Clinical Trials Registry: DRKS00023526). The trial was approved by the local ethics authorities (Reg no. 320_20B) and designed in cooperation with three official patient partners (GB, MK, CE; Deutsche Rheuma-Liga Bundesverband e.V). Participants were consecutively recruited at the outpatient clinic of the Department of Internal Medicine 3 (FAU Erlangen-Nurnberg) between November 2020 and February 2021. To be included, patients had to fulfill the EULAR/ACR classification criteria for RA [14] and had to have antibodies against cyclic citrullinated peptides (anti-CCP) or rheumatoid factor (RF IgM).

Participants were then randomly allocated in a 1:1 ratio to self-collection of the capillary blood via finger prick (BD MicrotainerTM) (FP) or by a device designed for capillary blood collection from the upper-arm (via Tasso-SSTTM device) (UA). We used simple randomization, akin to random draws without replacement from a closed urn containing the planned total number of assignments. The randomization list was concealed by a member of the study team, who only made treatment allocations as per sequence when requested and was not otherwise involved in the trial. Patients in the FP group used traditional blue BD MicrotainerTM lancets (Becton, Dickinson and Company, blade diameters 1.5 mm (W) × 2.0 mm (D), Catalog No 366594, Franklin Lakes, NJ, USA) and BD SSTTM Microtainer blood collection tubes (Becton, Dickinson and Company, Catalog No 365967, Franklin Lakes, NJ, USA), allowing collection of up to approximately 1800 μl. After twisting off the safety cap, the lancet is activated by pressing against the finger in the FP group. Subsequently, the user needs to massage the finger and collect the capillary blood into a separate collection tube. Patients in the UA group used a Tasso-SSTTM device (Tasso Inc., Seattle, WA, USA), which is a self-adhesive lancet-based device to be used on the upper arm instead of the fingertip (Fig. 1). The device is attached to the upper arm by an adhesive, and the lancet is activated by pressing a button. Upon skin puncture, the device applies a vacuum to increase capillary blood flow. The blood is automatically collected into the attached tube with a maximum capacity of approximately 500 μl. The researchers who recruited and screened participants for eligibility were kept blinded to the randomization status. All patients provided written informed consent to participate in the study.

Fig. 1
figure 1

Capillary blood self-sampling devices. Left: Tasso SSTTM device used for upper-arm (UA) capillary blood sampling; right: BD MicrotainerTM finger prick device used in the fingertip sampling group (FP)

Sample collection

Patients were instructed on how to use the self-sampling devices by local health care professionals and asked as to whether they had previously used self-sampling devices (for example for diabetes). In the UA group, written instructions were provided and a publicly available instruction video was presented by muting the original audio track in English, while the instructions displayed in the video were simultaneously explained in German. In the FP group, only oral instructions were given in German. Instructions included the sampling procedure, common pitfalls and how to avoid them. The primary instruction was to fill up the blood sampling containers within a maximum of 5 min. Patients were free to collect the samples from the dominant or nondominant arm. Assistance was available upon request, and study personnel recorded reasons for assistance. The time elapsed from skin puncture to retraction of blood collection device was recorded. Matched venous blood samples were obtained by trained phlebotomists from all participants within 1 h of capillary blood collection. The blood samples were kept at room temperature for at least 30 min, centrifuged for 15 min at 3200 RPM, and stored at 4 °C for 7–14 days. Samples were shipped to Thermo Fisher Scientific laboratory in Freiburg, Germany. Samples were inspected independently by two experienced lab technicians for quality. Upon arrival in the laboratory, the serum was transferred into Sarstedt™ 2 mL Polypropylene Micro Tubes (Sarstedt AG & Co., Nümbrecht, Germany) and stored at −20 °C until analysis. RF-IgM (EliA™ RF IgM, Thermo Fisher Scientific, Germany) and CCP-IgG antibodies (EliA™ CCP, Thermo Fisher Scientific, Freiburg, Germany) were analyzed on the PhadiaTM 250 instrument, and CRP was analyzed using Thermo Scientific™ Indiko™ Plus Clinical Chemistry Analyzer, Dreieich, Germany, once all samples were collected. While RF-IgM and CCP-IgG antibodies required only 5 μl of serum because the lab prepared manual dilutions, CRP measurement additionally required at least 100 μl of serum, so that CRP values were only available for samples with at least 105 μl of serum volume.

Outcome measures

The primary outcome was the agreement of RF-IgM, CCP-IgG antibody, and CRP levels between matched capillary and venous samples. Secondary outcomes included pain perception, usability, general acceptance, and number of attempts for sampling. Volume and duration of the sample collection were also assessed for UA, FP, and venous blood sampling (VBS). Pain perception of capillary blood sampling and venipuncture was measured using a numeric pain rating-scale (NRS; 0 no pain at all, 10 worst imaginable pain) [15] directly after blood collection. Capillary sample volume and sample acquisition time were recorded as the time between perforation of the skin and closing of collection tube (maximum of 5 min). The usability was assessed with the validated System Usability Scale (SUS) [16].

Based on the adjective SUS rating scale as described by Bangor et al. [17], these values were translated to categories such as “excellent.” Acceptability was assessed using the Net Promoter Score (NPS) [18]. The NPS [18] asks patients how likely they are to recommend something to a friend or patient. Participants answer using a 11-point numeric rating scale (0-not at all likely to 10-extremely likely). Answers between 0 and 6 are summarized as detractors, 7–8 as passives, and 9–10 as promoters. The NPS is calculated by subtracting the percentage of detractors from the percentage of promoters. Finally, we assessed the proportion of patients willing to apply self-sampling at home and stating that they clearly understood when the self-sampling process was finished using the respective device.

Statistical analysis

The sample size for this pilot study was based on convenience, and no formal sample-size calculation was undertaken. We summarized study group characteristics using appropriate summary statistics. We estimated intraclass correlation coefficients with 95% confidence intervals to quantify the agreement of analysis results obtained by venous and capillary sampling both overall and stratified by study groups. We also estimated and plotted Bland-Altman limits of agreement. Since the distributions of CCP, RF-IgM, and CRP tend to be lognormal, in addition to the conventional limits of agreement, we also estimated multiplicative limits of agreement using log differences that correspond to the ratio of values obtained from the venous sample and the capillary sample. Using the standard deviation of these log differences, limits of agreement were calculated separately for each mean value of capillary and venous sample results, based on the method described by Euser et. al. [19]. Mann-Whitney’s U test was used to compare continuous variables and scores for acceptability and pain in the two study arms and Wilcoxon’s signed-rank test for comparing pain scores between capillary and venous blood collection. Categorical variables were compared using the Fisher’s exact test. The level of significance was set as two-tailed p < 0.05 for all statistical tests. Analyses were carried out using the R software environment (version 4.0.1; R Foundation for Statistical Computing, Vienna, Austria).

Results

Participants

A total of 67 seropositive RA patients were screened for eligibility, and 50 patients were randomized (Fig. 2). Demographic and disease characteristics are summarized in Table 1.

Fig. 2
figure 2

Patient flow diagram. Depiction of the number of screened rheumatoid arthritis patients, the number and reasons for screening failure, and the number of patients randomized to the two groups

Table 1 Patient demographics and disease characteristics

Capillary blood sample collection quality and procedure

In total, 49 matched capillary and venous samples were obtained, providing enough volume for the analysis of CCP and RF in 49 participants and additionally CRP in 22 participants (Figure S1). In the UA group, one patient was not able to collect blood despite attempting with a second device. In both groups, 4/25 (16%) of the patients were able to collect samples only in a second attempt (using a new self-sampling device) due to initial failure. In UA and FP groups, 8/25 (32%) and 7/25 (28%), respectively, requested assistance from the study personnel to perform the blood collection. The mean blood volume collected for UA and FP was 106.2 ± 60.6 μl and 118.8 ± 74.7 μl, respectively. The mean blood collection time for UA and FP was 256.8 ± 79.1 s and 230.6 ± 78.3 s, respectively.

Agreement of anti-CCP Ab, RF-IgM, and CRP results between capillary and venous blood samples

We observed excellent agreement between capillary and venous analyses. Intraclass correlation coefficients (95% CI) were high for anti-CCP (0.968; 0.937 to 0.984), RF (0.993; 0.986 to 0.996), and CRP (0.998; 0.996 to 0.999) in the UA group (Fig. 3 and Figure S2). Intraclass correlation coefficients (95% CI) were also high for anti-CCP (0.998; 0.995 to 0.999), RF (0.996; 0.991 to 0.998), and CRP (0.992; 0.984 to 0.996), in the FP group (Fig. 3 and Figure S2). The Bland-Altmann limits of agreement are shown in Fig. 4.

Fig. 3
figure 3

Agreement between capillary and venous blood sampling with respect to C-reactive protein and autoantibody results. Dots show the intraclass correlation coefficients, lines the 95% confidence intervals between self-sampling of the capillary blood from the upper arm (red) or via finger pricking (green) and venous blood sampling. Blue bars show the combined results of the upper arm and finger prick sampling. CCP, anti-cyclic citrullinated peptide antibodies; RF, rheumatoid factor immunoglobulin M; CRP, C-reactive protein

Fig. 4
figure 4

Bland-Altman comparison of the capillary blood and venous blood results with respect to C-reactive protein and autoantibody results. Bland-Altman diagrams showing differences in the measurements of anti-cyclic citrullinated peptide antibodies (CCP; U/ml), rheumatoid factor immunoglobulin M (RF; IU/ml)), and C-reactive protein (CRP; mg/l) between venous blood sampling and capillary blood sampling (upper arm, left; finger pricking, middle; combined groups, right). The dotted lines represent the ideal mean difference, the blue lines represent the observed mean difference, and the red and green lines indicate the limits of agreement on the additive scale and multiplicative scale, respectively

Usability, pain, and acceptance of self-sampling

In the UA sampling group, 17/25 (68%) of the participants and in the FP sampling group 21/25 (84%) of the participants (p=0.32 by exact test) “agreed” or “completely agreed” that the respective device was easy to use. Mean SUS scores were 83.1 and 80.7 for UA and FP, respectively, translating to “excellent” usability using the method described by Bangor et al. [17] (Table 2). We did not observe substantial differences between the groups in individual item or total scores. Furthermore, 24/25 (96%) of patients in the UA group and 25/25 (100%) in the FP group stated that the instructions were clear. The end to blood collection was clearly discernible for 16/25 (64%) participants in the UA group and 19/25 (76%) in the FP group.

Table 2 Means and standard deviation scores for the System Usability Scale

The mean pain NRS (range 0–10) scores for capillary self-sampling (UA: 1.7 ± 1.1; FP: 1.9 ± 1.9, p=0.93) were similar between the groups (p=0.93) and significantly lower than that in the standard venous blood collection (2.8 ± 1.7 in UA and 2.1 ± 2.0 in the FP group, p=0.003 overall). Numerically more patients in the UA group compared to the FP group experienced self-sampling as less painful compared to standard venous blood collection, 15/25 (60%) vs 9/25 (36%), see Fig. 5. The proportion of promoters was 11/25 (44%) in the UA group vs 8/25 (32%) in the FP group while 4/25 (16%) vs 13/25 (52%) were detractors, respectively (exact p=0.016 for all categories), resulting in a positive NPS for the UA group of +28% vs a negative NPS of −20% for the FP group, see Fig. 5. 15/25 (60%) and 8/25 (32%) stated that they would like to independently collect the capillary blood at home instead of seeing a professional for a venous blood collection in the UA and FP group, respectively.

Fig. 5
figure 5

A Percentage of patients per group for respective change in pain and B percentage of patients per group for respective promoter score category

Discussion

This study showed excellent agreement of RF IgM, anti-CCP IgG, and CRP measurements between the capillary blood obtained from self-sampling and venous blood obtained by health care professionals in RA patients. The usability of both self-sampling devices was rated excellent by the participants. Most importantly, patients experienced less pain compared to traditional venous blood collection and upper-arm based self-sampling appears to be preferred over finger pricking by RA patients.

Laboratory results are essential to monitor disease activity of RA and to allow early detection of the disease [20]. Adding patient-centered self-sampling of the blood to already existing digital symptom self-assessment devices could therefore significantly increase the potential to monitor RA patients in a so far unprecedented way [7, 21, 22]. Furthermore, it can be expected that self-sampling could empower patients allowing them to time-independently obtain laboratory results that in case of CRP levels allow to objectify disease activity and potentially also to predict flares [23, 24]. While this approach is new to rheumatology, Pedersen and colleagues reported that self-monitoring (weekly questionnaire completion + fecal calprotectin test) using a web application helped to personalize treatment for patients suffering from ulcerative colitis [25] and Crohn’s disease [26]. In a recent feasibility trial patient self-sampling helped to substantially improve urate levels in patients with gout [27].

Our results are supported by a recent study that evaluated the perception, painfulness, and usability of the fingertip versus upper arm capillary blood collection among national athletes from Denmark [12]. Although another Tasso upper-arm device was used in the study, the self-sampling procedure and withdrawal mechanism are highly comparable. The associated pain using the upper arm device was rated lower compared to finger pricking (−0.4 ± 1.6, p < 0.05). Interestingly, the large majority (96%) of the athletes preferred capillary-based blood collection over venous blood collections and the majority (78%) preferred collection from the upper arm over the fingertip. In addition, also Blicharz and colleagues reported a significantly lower pain score associated with a similar upper arm device (TAP; Seventh Sense Biosystems, Medford, USA) compared to venous blood collection (0.4 vs 1.5 on pain scale from 0 to 10) investigating 143 participants [13]. A recent study evaluating the same upper-arm device also showed high correlation between venous and capillary blood when testing for anti-SARS-CoV-2 antibodies [28].

The upper-arm device failure rate was higher in our study compared to previous studies performed with the same device (20% (6/30) vs 2% (2/108) [12] and 4% (10/240) [29]). A main reason for this difference could be the fact that capillary blood collection in our study was carried out by patients themselves and not by healthy individuals or health care professionals. Our study collective had RA and therefore was to some extent handicapped by the disease. Indeed, self-collection of the blood in the UA group was impaired as patients were requesting help to press the button (5/25; 20%) and to remove the blood collection container (3/25; 12%). Using the same upper-arm device, healthy participants were able to successfully draw blood more often during the first draw and without assistance in a previous study [28] (first draw: 93 vs 80%; second draw 7 vs 16%). Future studies should analyze the failure rate longitudinally, as experienced users might use these devices more successfully. Additional to this aspect, we are currently prospectively investigating self-sampling feasibility in different rheumatic patient populations, at home scenarios and healthy individuals, also including qualitative patient feedback.

Similar to the aforementioned athlete study, RA patients did attest UA sampling a higher usability (SUS) than FP sampling. However, when plotting our SUS results on an adjective rating scale as described by Bangor et al. [17], SUS scores for both devices were rated “excellent”. Also, acceptance was higher for the UA than for FP sampling (NPS +28% vs. −20). Furthermore, the majority of UA patients (60% compared to 32% in FP group) stated that they would like to independently collect capillary blood at home instead of seeing a healthy professional for a venous blood collection. Offering patients a variety of blood collection sites is likely to decrease pain and discomfort. Regarding patients with veins difficult to detect even for health care professionals, patients with fear of needles, bruising, and general anxiety, capillary-based self-sampling of the blood may be a good alternative.

The parallel group design is a limitation of this study. Patients could have used both the UA and the FP device sequentially allowing a direct comparison; however, the sequence of device use may have also caused priming and bias in patient-reported outcomes. We have therefore deliberately chosen a parallel group design, which has also allowed us to reduce the number of blood collections. Furthermore, the limited volume of the blood obtained by the capillary collection devices impeded the analysis of CRP in a number of patients. This is a limitation of capillary blood sampling requiring further development in device design. Simulation of a an “at-home” situation and manual work needed to analyze probes are further limitations of this study. On the other hand, patient involvement in the study design and in conducting the study represents strengths of this study. In addition, the capillary self-sampling devices were used by patients themselves and not by experienced health care professionals, as in previous studies [12, 13, 29].

To our knowledge, this is the first rheumatology study that has performed a structured analysis on self-sampling of the capillary blood for the measurement of inflammation markers and autoantibodies. Accuracy, feasibility, and acceptance of the two self-sampling devices were high providing the possibility for remote analysis. While more studies are needed to effectively implement this novel technology in disease monitoring, our data provide evidence for the principal feasibility of such an approach. Considering the higher acceptance of UA sampling, future research should focus on analyzing such devices, i.e., the standardization of self-sampling procedures such as application of chemical heat pads [29] to improve blood circulation in the skin and blood volume output. Furthermore, online counseling [30, 31] could guarantee the assistance that was necessary in some of the patients.

Conclusion

The excellent usability and high concordance with the results from venous blood analysis illustrate the potential of capillary blood self-sampling. This approach does not only provide more comfort, higher flexibility, and less pain but also supports tight disease monitoring and potentially also improves the early recognition of RA.

Availability of data and materials

Data analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ACR:

American College of Rheumatology

CCP:

Cyclic citrullinated peptides

CRP:

C-reactive Protein

EULAR:

European alliance of associations for rheumatology

FP:

Finger prick

NPS:

Net Promoter Score

NRS:

Numeric rating scale

RA:

Rheumatoid arthritis

RF:

Rheumatoid factor

SUS:

System Usability Scale

UA:

Upper-arm

VBS:

Venous blood sampling

References

  1. McInnes IB, Schett G. Pathogenetic insights from the treatment of rheumatoid arthritis. Lancet. 2017;389:2328–37.

    Article  CAS  Google Scholar 

  2. Krusche M, Sewerin P, Kleyer A, Mucke J, Vossen D, Morf H, et al. Specialist training quo vadis? Z Rheumatol. 2019;78:692–7.

    Article  CAS  Google Scholar 

  3. Dejaco C, Alunno A, Bijlsma JW, Boonen A, Combe B, Finckh A, et al. Influence of COVID-19 pandemic on decisions for the management of people with inflammatory rheumatic and musculoskeletal diseases: a survey among EULAR countries. Ann Rheum Dis. 2020;annrheumdis-2020-218697.

  4. Burmester GR, Pope JE. Novel treatment strategies in rheumatoid arthritis. Lancet. 2017;389:2338–48.

    Article  Google Scholar 

  5. Kernder A, Morf H, Klemm P, Vossen D, Haase I, Mucke J, et al. Digital rheumatology in the era of COVID-19: results of a national patient and physician survey RMD Open. BMJ Specialist J. 2021;7:e001548.

    Google Scholar 

  6. Morf H, Krusche M, Knitza J. Patient self-sampling: a cornerstone of future rheumatology care? Rheumatol Int. 2021;41:1187–8.

    Article  Google Scholar 

  7. Knitza J, Knevel R, Raza K, Bruce T, Eimer E, Gehring I, et al. Toward earlier diagnosis using combined eHealth tools in rheumatology: the Joint Pain Assessment Scoring Tool (JPAST) Project. JMIR Mhealth Uhealth. 2020;8:e17507.

    Article  Google Scholar 

  8. Dobreva T, Brown D, Park JH, Thomson M. Single cell profiling of capillary blood enables out of clinic human immunity studies. Sci Rep. 2020;10:20540.

    Article  CAS  Google Scholar 

  9. Bateman P, K. The development of patient-centric sampling as an enabling technology for clinical trials. Bioanalysis. 2020;12:971–6.

    Article  Google Scholar 

  10. James CA, Barfield MD, Maass KF, Patel SR, Anderson MD. Will patient-centric sampling become the norm for clinical trials after COVID-19? Nat Med. 2020;26:1810.

    Article  CAS  Google Scholar 

  11. Qu Y, Brady K, Apilado R, O’Malley T, Reddy S, Chitkara P, et al. Capillary blood collected on volumetric absorptive microsampling (VAMS) device for monitoring hydroxychloroquine in rheumatoid arthritis patients. J Pharm Biomed Anal. 2017;140:334–41.

    Article  CAS  Google Scholar 

  12. Solheim SA, Ringsted TK, Nordsborg NB, Dehnes Y, Levernaes MCS, Mørkeberg J. No pain, just gain: painless, easy, and fast dried blood spot collection from fingertip and upper arm in doping control. Drug Test Anal. 2021.

  13. Blicharz TM, Gong P, Bunner BM, Chu LL, Leonard KM, Wakefield JA, et al. Microneedle-based device for the one-step painless collection of capillary blood samples. Nat Biomed Eng. 2018;2:151–7.

    Article  CAS  Google Scholar 

  14. 2010 Rheumatoid arthritis classification criteria: an American College of Rheumatology/European League Against Rheumatism collaborative initiative - Aletaha - 2010 - Arthritis & Rheumatism - Wiley Online Library [Internet]. [cited 2021 Jul 15]. Available from: https://onlinelibrary.wiley.com/doi/full/10.1002/art.27584

  15. Karcioglu O, Topacoglu H, Dikme O, Dikme O. A systematic review of the pain scales in adults: Which to use? Am J Emerg Med. 2018;36:707–14.

    Article  Google Scholar 

  16. Bangor A, Kortum PT, Miller JT. An empirical evaluation of the system usability scale. International Journal of Human–Computer Interaction. Taylor & Francis; 2008;24:574–594.

  17. Bangor A, Kortum P, Miller J. Determining what individual SUS scores mean: adding an adjective rating scale. J Usability Studies. 2009;4:114–23.

    Google Scholar 

  18. Reichheld F. The one number you need to grow [Internet]. Harv Bus Rev. 2003; Available from: https://hbr.org/2003/12/the-one-number-you-need-to-grow.

  19. Euser AM, Dekker FW, le Cessie S. A practical approach to Bland-Altman plots and variation coefficients for log transformed variables. J Clin Epidemiol. 2008;61:978–82.

    Article  Google Scholar 

  20. Ehrenstein B, Pongratz G, Fleck M, Hartung W. The ability of rheumatologists blinded to prior workup to diagnose rheumatoid arthritis only by clinical assessment: a cross-sectional study. Rheumatology (Oxford). 2018;57:1592–601.

    Article  Google Scholar 

  21. Knitza J, Mohn J, Bergmann C, Kampylafka E, Hagen M, Bohr D, et al. Accuracy, patient-perceived usability, and acceptance of two symptom checkers (Ada and Rheport) in rheumatology: interim results from a randomized controlled crossover trial. Arthritis Res Therapy. 2021;23:112.

    Article  CAS  Google Scholar 

  22. Knitza J, Tascilar K, Gruber E, Kaletta H, Hagen M, Liphardt A-M, et al. Accuracy and usability of a diagnostic decision support system in the diagnosis of three representative rheumatic diseases: a randomized controlled trial among medical students. Arthritis Res Therapy. 2021;23:233.

    Article  Google Scholar 

  23. Trouvin A-P, Jacquot S, Grigioni S, Curis E, Dedreux I, Roucheux A, et al. Usefulness of monitoring of B cell depletion in rituximab-treated rheumatoid arthritis patients in order to predict clinical relapse: a prospective observational study. Clin Exp Immunol. 2015;180:11–8.

    Article  CAS  Google Scholar 

  24. Orange DE, Yao V, Sawicka K, Fak J, Frank MO, Parveen S, et al. RNA identification of PRIME cells predicting rheumatoid arthritis flares. N Engl J Med; 2020 [cited 2021 Oct 15]; Available from: https://doi.org/10.1056/NEJMoa2004114

  25. Pedersen N, Thielsen P, Martinsen L, Bennedsen M, Haaber A, Langholz E, et al. eHealth: individualization of mesalazine treatment through a self-managed web-based solution in mild-to-moderate ulcerative colitis. Inflamm Bowel Dis. 2014;20:2276–85.

    Article  Google Scholar 

  26. Pedersen N, Elkjaer M, Duricova D, Burisch J, Dobrzanski C, Andersen NN, et al. eHealth: individualisation of infliximab treatment and disease course via a self-managed web-based solution in Crohn’s disease. Aliment Pharmacol Ther. 2012;36:840–9.

    Article  CAS  Google Scholar 

  27. Riches PL, Alexander D, Hauser B, Kuske B, Krause A. Evaluation of supported self-management in gout (GoutSMART): a randomised controlled feasibility trial. Lancet Rheumatol. Elsevier. 2022;4:e320–8.

    Article  Google Scholar 

  28. Hendelman T, Chaudhary A, LeClair AC, van Leuven K, Chee J, Fink SL, et al. Self-collection of capillary blood using Tasso-SST devices for Anti-SARS-CoV-2 IgG antibody testing. PLoS One. 2021;16:e0255841.

    Article  CAS  Google Scholar 

  29. Roadcap B, Hussain A, Dreyer D, Carter K, Dube N, Xu Y, et al. Clinical application of volumetric absorptive microsampling to the gefapixant development program. Bioanalysis. Future Science. 2020;12:893–904.

    Article  CAS  Google Scholar 

  30. Valentine-Graves M, Hall E, Guest JL, Adam E, Valencia R, Shinn K, et al. At-home self-collection of saliva, oropharyngeal swabs and dried blood spots for SARS-CoV-2 diagnosis and serology: post-collection acceptability of specimen collection process and patient confidence in specimens. PLoS One. 2020;15:e0236775.

    Article  CAS  Google Scholar 

  31. Wang Z, Lau JTF, Ip M, Ho SPY, Mo PKH, Latkin C, et al. A randomized controlled trial evaluating efficacy of promoting a home-based HIV self-testing with online counseling on increasing HIV testing among men who have sex with men. AIDS Behav. 2018;22:190–201.

    Article  Google Scholar 

Download references

Acknowledgements

We thank all patients for their participation in this study. This study is part of the PhD thesis of the first author JK (AGEIS, Université Grenoble Alpes, Grenoble, France). We thank Josefine Born and Deniz Krämer for their help recruiting patients.

Funding

Open Access funding enabled and organized by Projekt DEAL. The study was supported by the Deutsche Forschungsgemeinschaft (DFG – FOR 2886 “PANDORA” - B01/A03/Z01 to G.K. G.S. and A.K.) and Thermo Fisher Scientific (Freiburg, Germany).

Author information

Authors and Affiliations

Authors

Contributions

All authors have made substantial contributions to the conception, design of the work, the analysis, and interpretation of data. All authors contributed to drafting of the work and substantively revised it. The authors read and approved the final manuscript.

Corresponding author

Correspondence to Johannes Knitza.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the regional ethics review board in Erlangen, Germany (Reg no. 320_20B). All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual respondents included in the study.

Consent for publication

All authors of the manuscript have read and agreed to its content and are accountable for all aspects of the accuracy and integrity of the manuscript in accordance with ICMJE criteria.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Fig. S1.

Self-sampling feasibility results according to randomization arm

Additional file 2: Fig. S2.

Intraclass correlation coefficients and 95% confidence interval by group and analyte

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Knitza, J., Tascilar, K., Vuillerme, N. et al. Accuracy and tolerability of self-sampling of capillary blood for analysis of inflammation and autoantibodies in rheumatoid arthritis patients—results from a randomized controlled trial. Arthritis Res Ther 24, 125 (2022). https://doi.org/10.1186/s13075-022-02809-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13075-022-02809-7

Keywords